39
Alarm categories
The alarms are classified into four categories according to the priority depends primarily on the cause and alarm duration. The priority
increasing with the duration and according to the physiological significance.
NOTE: Asystole, ventricular fibrillation and V Tach alarms are always high priority alarms.
For more information about audible tone pattern see the “Default Configuration Worksheet”.
Signal
Priority level
High
Medium
Low
Informational
Alarms in Message Field Solid red background
Yellow boundary, blue-
gray background
Cyan boundary, blue-
gray background
Blue-gray background
field
Alarm reset in Message
Field
Solid red background
with audio pause symbol
Yellow boundary, blue-
gray background, with
audio pause symbol
Cyan boundary, blue-
gray background, with
audio pause symbol
No
Physiological data
values in Digit Field
Flashing red,
physiological alarms only
Flashing yellow,
physiological alarms only
No
No
Alarm light
Flashing red
Flashing yellow
Solid Cyan
No
Alarm tone ISO pattern
10 beeps every 5 seconds 3 beeps every 19 seconds 1 beep every 30 seconds No
Alarm tone ISO2 pattern 10 beeps every 5 seconds
(rising tone)
3 beeps every 19.5
seconds (rising tone)
1 beep every 30 seconds No
Alarm tone IEC pattern
10 beeps every 5 seconds
(rising tone)
3 beeps every 19 seconds
(rising tone)
1 beep every 30 seconds No
Alarm tone Gerenal
pattern
Continuous beep
Double beep every 5
seconds
1 beep every 30 seconds No