142
Nellcor SpO
2
specifications
Masimo SpO
2
specifications
Measurement range
1 to 100%
Display range
0 to 100%
Calibrated against functional oxygen saturation.
Measurement accuracy
Adult 100 to 70% ±2 digits
Neo 100 to 70% ±3 digits
Low perfusion 100 to 70% ±2 digits
Display resolution
1% of SpO
2
Display averaging
2 to 7 seconds
Pulse rate
Measurement and display
range
20 to 250 bpm
Display resolution
1 bpm
Measurement accuracy
±3 digits
Sensor Light Source*
Wavelength
Infrared: 900 nm
Red: 660 nm
Total optical output power
of the sensor LEDs
less than 15 mW
*
This information may be useful to clinicians, such as those performing
photodynamic therapy.
Measurement range
1 to 100%
Display range
0 to 100%
Calibrated against functional oxygen saturation.
Measurement accuracy
Without motion
Adult/Pediatric 100 to 70% ±2 digits
Neonate 100 to 70% ±3 digits
With motion
Adult/Ped/Neo 100 to 70% ±3 digits
Low perfusion
100 to 70% ±2 digits
0~69% unspecified
Display resolution
1% of SpO
2
Display averaging
2 to 16 seconds
Pulse rate
Measurement and display
range
25 to 240 bpm
Display resolution
1 bpm
Measurement accuracy
Without motion ±3 bpm
With motion ±5 bpm
Sensor Light Source*
Wavelength
Infrared: 905 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation
Infrared: 22.5 mW (max)
Red: 27.5 mW (max)