Using the manual Entropy sensor check
Whenever required, you can perform the sensor check manually.
1. Press the
Check Sensor
module key, or:
−
Press the
Entropy
module key, or press the
Others
key,
select
Entropy
−
Select
Check Sensor.
2. Observe the results on the screen. Do not press the sensor
during the check to avoid signal noise.
3. The measurement continues automatically after the sensor has
passed the check.
Bypassing the Entropy sensor check
If the sensor does not pass the impedance check, this option
becomes selectable. It allows you to start the measurement without
completing the sensor check.
NOTE: In this case, the measurement may be unreliable.
1. Press the
Entropy
module key, or press the
Others
key, select
Entropy
2. Select
Bypass Check
.
Using the automatic Entropy sensor check
To take the automatic sensor check into use, the monitor will check
the sensor every 10 minutes.
1. Press the
Entropy
module key, or press the
Others
key, select
Entropy
2. Select
Automatic Check
to ON/OFF.
NOTE: This menu doesn’t influence the initial check, the system will
automatically do the self check once patient connect.