110
O
2
measurement
NOTE: B40 only.
Selecting the O
2
scale
If the difference between FiO
2
and EtO
2
is above 6%, change the O
2
scale:
1. Press the
Airway Gas
key.
2. Select
O2 Setup
.
3. Select
Scale
4. Select an option from the scale list.
OFF/ON the O
2
measurement
1. Press the
Airway Gas
key.
2. Select
O2 Setup
-
Measurement
.
3. Choose
OFF
or
ON
Setting O
2
alarms
1. Press the
Airway Gas
key.
2. Select
O2 Setup
-
O2 Alarm
.
3. Select
Adjust Limits
to set up the limits.
4. Select
EtO2 Alarm, FiO2 Alarm
to OFF/ON the limits.
5. You can also select
Alarms Setup
to adjust O
2
or related alarms.
Anesthetic agent and N
2
O measurement
NOTE: B40 only.
NOTE: The E-sCAiO and N-CAiO modules automatically identify the
agent being used. If the agent value is out of measurment range,
“---” will be displayed on screen, which means the value is invalid.
Selecting the agent scale
1. Press the
Airway Gas
key.
2. Select
Agent/N2O Setup
-
Agent Scale
3. Select an option from the scale list.
OFF/ON the agent measurement
1. Press the
Airway Gas
key.
2. Select
Agent/N2O Setup
-
Agent Measurement
.
3. Choose
OFF
or
ON
OFF/ON the N
2
O measurement
1. Press the
Airway Gas
key.
2. Select
Agent/N2O Setup
-
N2O Measurement
.
3. Choose
OFF
or
ON