Site Preparations
2-10
Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
Probes environmental requirements
Operation and storage temperatures for probes
NOTE:
Systems and electronic probes are designed for storage
temperatures of -5 to +50°C. When exposed to large
temperature variations, the product should be kept in room
temperature for 10 hours before use.
Table 2-7: Operation and storage temperatures for probes
Conditions
Temperature
Operation:
10 to 40°C (50 to 104°F)
Storage:
-5 to 50°C (23 to 122°F)
Temperature in degrees Celsius (°C) conversion to degrees F (°F):
(°F) = (°C * 9/5) + 32
Summary of Contents for 5448623
Page 5: ...Venue 50 Basic Service Manual i 3 5447566 100 English Rev 7 ...
Page 6: ...i 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 7: ...Venue 50 Basic Service Manual i 5 5447566 100 English Rev 7 ...
Page 8: ...i 6 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 9: ...Venue 50 Basic Service Manual i 7 5447566 100 English Rev 7 ...
Page 10: ...i 8 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 11: ...Venue 50 Basic Service Manual i 9 5447566 100 English Rev 7 ...
Page 12: ...i 10 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 22: ...i 20 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 52: ...Introduction 1 30 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 72: ...Site Preparations 2 20 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 160: ...Components and Functions Theory 5 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 204: ...Troubleshooting 7 38 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 228: ...Replacement Procedures 8 24 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 284: ...Care and Maintenance 10 36 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 288: ...Index 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 289: ......