Electrical safety tests
Venue 50 – Basic Service Manual
10-25
5447566-100 English Rev. 7
Grounding continuity
Measure the resistance from the third pin of the attachment plug
to the exposed metal parts of the case. The ground wire
resistance should be less than
0.1
ohms. Reference the
procedure in the IEC60601-1.
Figure 10-3. Ground continuity test
DANGER
ELECTRIC SHOCK HAZARD. THE PATIENT MUST NOT BE
CONTACTED TO THE EQUIPMENT DURING THIS TEST.
1. GROUND
PIN
2. OHMMETER
3. DOCKING
CART
4. ACCESSIBLE METAL PART:
• I/O Board
• Probe cable hook
Summary of Contents for 5448623
Page 5: ...Venue 50 Basic Service Manual i 3 5447566 100 English Rev 7 ...
Page 6: ...i 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 7: ...Venue 50 Basic Service Manual i 5 5447566 100 English Rev 7 ...
Page 8: ...i 6 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 9: ...Venue 50 Basic Service Manual i 7 5447566 100 English Rev 7 ...
Page 10: ...i 8 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 11: ...Venue 50 Basic Service Manual i 9 5447566 100 English Rev 7 ...
Page 12: ...i 10 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 22: ...i 20 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 52: ...Introduction 1 30 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 72: ...Site Preparations 2 20 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 160: ...Components and Functions Theory 5 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 204: ...Troubleshooting 7 38 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 228: ...Replacement Procedures 8 24 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 284: ...Care and Maintenance 10 36 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 288: ...Index 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 289: ......