GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
10-8
Section 10-5 - System Maintenance
10-5-5
Physical Inspection
Table 10-9 Physical Checks
Step
Item
Description
1
Labeling
Verify that all system labeling is present and in readable condition. Refer to User Manual,...... for
details.
2
Scratches & Dents Inspect the console for dents, scratches or cracks.
3
Control Panel
Inspect keyboard and control panel. Note any damaged or missing items.
4
Cables &
Connectors
Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
special attention to footswitch assembly and probe strain or bend reliefs.
5
Shielding &
Covers
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Missing covers and hardware could cause EMI/RFI problems while scanning.
6
External I/O
Check all connectors for damage.
7
Op Panel Lights
Check for proper operation of all operator panel and Freeze Key light.
Summary of Contents for 5151219