GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
4-12
Section 4-3 - General Procedure
4-3-7-3
B Mode Softmenu Key
Table 4-4
B Mode Softmenu Key
Step
Task
Expected Result(s)
comments
1
Adjust Rejection
Selects a level below which echoes will not be
amplified (an echo must have a certain
minimum amplitude before it will be processed).
2
Activate Colorize
Enables gray scale image colorization. To
deactivate, reselect a Gray Map.
3
Adjust Edge Enhance
Edge Enhance brings out subtle tissue
differences and boundaries by enhancing the
gray scale differences corresponding to the
edges of structures. Adjustments to M Mode's
edge enhancement affects the M Mode only.
For LOGIQ e, LOGIQ e Vet and
LOGIQ i
Adjust Edge Enhance/Contour
Edge Enhance brings out subtle tissue
differences and boundaries by enhancing the
gray scale differences corresponding to the
edges of structures. Adjustments to M Mode's
edge enhancement affects the M Mode only.
For Vivid e
4
Activate Gray Map
Determines how the echo intensity levels
received are presented as shades of gray.
5
Adjust Frequency
Multi Frequency mode lets you downshift to the
probe's next lower frequency or shift up to a
higher frequency.
6
Adjust Frame Average
Temporal filter that averages frames together.
This has the effect of presenting a smoother,
softer image.
For LOGIQ e, LOGIQ e Vet and
LOGIQ i
Adjust DDP
Temporal filter that averages frames together.
This has the effect of presenting a smoother,
softer image.
For Vivid e
7
Adjust Rotation
Rotates the image by selecting the value from
the pop-up menu.
8
Adjust Line Density
Optimizes B Mode frame rate or spatial
resolution for the best possible image.
For LOGIQ e, LOGIQ e Vet and
LOGIQ i
Adjust Frame Rate
Optimizes B Mode frame rate or spatial
resolution for the best possible image.
For Vivid e
9
Power output
Optimizes image quality and allows user to
reduce beam intensity. 2% increments between
0-100%. Values greater than 0.1 are displayed.
10
Dynamic Range
Dynamic Range controls how echo intensities
are converted to shades of gray, thereby
increasing the adjustable range of contrast.
11
Focus Number and Position
Increases the number of transmit focal zones or
moves the focal zone(s) so that you can tighten
up the beam for a specific area. A graphic caret
corresponding to the focal zone position(s)
appears on the right edge of the image.
12
Virtual Convex
Virtual Convex for linear probe
13
Virtual Apex
Virtual Convex for Sector probe
Only for LOGIQ e R5.x.x/R6.x.x/
R7.x.x, LOGIQ e Vet, LOGIQ i R5.x.x,
Vivid e R5.x.x, Vivid e R6.x.x
Summary of Contents for 5151219