
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 8 Replacement Procedures
8-11
8-3-2 Loading the System Software with USB Memory Stick
6.) The system software load instruction window displays. Select one of the options for loading
the system. Select choice [a] to load the complete disk. Refer to
-
To select [a], the complete disk will be loaded. This option is recommended for
applicationsoftware upgrade.
NOTE:
NOTE: When to select [a] to load complete disk, please ensure that any patient data on the disk
has been backed up.
-
To select [b], only the bootable C: partition is loaded. This option is intended for recovery of a
system that will not boot up. It is not recommended for application software upgrade because
during upgrade process, the data on the system would possibly be impacted.
-
To select [c] to quit system upgrade process
NOTICE
If you select [a], ALL existing software and data will be erased. If backup has not been
performed, all data like Patient Database, System Configuration and User Configurations
(Customer Presets) will be lost.
Figure 8-14 System Software load instruction
WARNING
WARNING
While the software install procedure is designed to preserve data, you should select
choice [b] to format disk C only.
Summary of Contents for 5151219
Page 271: ......