The design of the flow cell prevents the formation of distinct interfaces between eluent
components with different refractive indices and eliminates the negative influence these
would cause. The precision of monitoring is enhanced by the construction of the flow
cell, which ensures total reflectance of light. This maintains a high intensity of light to
the detector. The long path length combined with a small cell volume increases sensitiv-
ity.
Optional flow cells
The following flow cells can also be used with Monitor UV-900:
•
UV cell, 1/2/5 mm (ÄKTApilot)
•
Industrial flow cell 8 mm (tubing i.d. 8 mm)
•
Industrial flow cell 1” (tubing i.d. 1”)
26
Monitor UV-900 Operating Instructions 28962214 AD
3 System description
3.2 Flow cell