Tubing
Fixing UV Flow Cell 2 mm and UV
Flow Cell 10 mm
NOTICE
To avoid damaging the optical fibers, press only on the cell body,
never on the optical fibers.
Action
Step
Unpack the flow cell. Do not remove the red protective caps from the inlet
or outlet, or the black protective caps on the optical fiber connectors.
1
Slide the rear white clip on the cell holder to its inner position for a 2 mm
cell and to its outer position for a 10 mm cell.
2
32
Monitor UV-900 Operating Instructions 28962214 AD
4 Installation
4.4 Assembly