8.5 kg
Weight
IP 20
Degree of protection
< 80 dB A
Noise emission
UV flow cell, 2 and 10 mm
100 mL/min
Recommended maximum flow
rate
2 MPa (20 bar, 290 psi)
Maximum pressure
Maximum 0.5 bar at 2 mL/min with water at 25°C
Backpressure
4°C to 40°C
Liquid temperature range
2 mm
Optical path length, 2 mm cell
10 mm
Optical path length, 10 mm cell
2 μL
Cell volume, 2 mm cell
8 μL
Cell volume, 10 mm cell
PTFE (polytetrafluoroethylene)
Wetted materials
PEEK (polyetheretherketone)
Titanium (palladium alloy)
Quartz (synthetic fused silica)
1–13, 13–14 (< 1 days exposure)
pH stability range
The wetted parts are resistant to organic sol-
vents and salt buffers commonly used in chro-
matography of biomolecules, except 100% Ethyl
acetate, 100% Hexane, and 100% Tetrahydrofu-
ran (THF).
Chemical resistance
UNF 10-32 fingertight connectors for capillary
tubing with 1/16” outer diameter
Tubing connections
68
Monitor UV-900 Operating Instructions 28962214 AD
8 Reference information
8.1 Technical specifications