Revision 1.1
GM ECx Instruction Manual
9
Figure 3-1: GM ECx Explosion-Proof Housing
Modular design simplifies the installation of the GM ECx (Figure 3-2). A top Display Assembly is
mounted with captive thumbscrews and is easily removed to access internal components. The
RF900 module mounts
piggyback
to the back of the Display Assembly. Wiring from toxic or
oxygen sensors terminates at the10-0232 Display Assembly. This Display Assembly is the only
PC board supplied with the toxic / oxygen GM ECx. The enclosure is equipped with two threaded
3/4 inch NPT conduit fitting outlet and pre-drilled mounting flanges.
WARNING:
Qualified personnel should perform the installation according to applicable electrical
codes, regulations and safety standards. Insure correct cabling and sealing fitting practices are
implemented. Do not aim the sensor pointing upward. Install the GM ECx to a wall or bracket
using the predrilled mounting flanges with I.D. 0.25 on 5.5 inch centers (Figure 3-1). If conduit is
rigid and able to support the weight of the GM ECX, the mounting bolts may be omitted.
CAUTION: The sensor should never be installed pointing upwards
.
3.3.1 10-0322 Magnetic Mount Option
GDS Corp offers a magnetic mount for the GM ECx instrument enclosure
. The 10-0322Magnetic
Mount securely attaches the assembly to any solid steel structures that is at least 6 inches wide.
Figure 3-2: Outline Drawing