GASMAX ECx Instruction Manual
Revision 1.1
20
6.5 Sensor
Information:
Sensor Information
has the
SENSOR SETUP/INFO
menus shown in Figure 6-5
SENSOR SETTINGS/INFO
Install Sensor
Recall Cal History
Input Type SMART
New Sensor 01/28/04
Recent Cal 04/05/04
>
>
INSTALL NEW SENSOR
EDIT to install
New Sensor.
Any other key
to Exit.
CALIBRATION DATES
Record 22 of 22
CAL Date: 04/05/04
UP/DOWN to scroll
NEXT to EXIT.
INSTALL NEW SENSOR
Historical CAL data
Will be Deleted.
EDIT to Accept
NEXT to Abort
EDIT
EDIT
Sensor Life
>
SMART SENSOR
Type: Cat-Bead
Span: 100
Zero: 0
SN: xxxxxx
Born On: 01/28/04
Last Cal: 04/05/04
ANY key to Exit
Figure 6-5: Sensor Information Menus
Install New Sensor
should always be performed when a new
simple
sensor is installed. This
deletes historical CAL data and sets sensor life to 100% after initial calibration of the new
simple
sensor. The GM ECX
Smart
sensor interface will automatically detect new smart sensors and
this menu is therefore not available with a smart sensor connected.
Recall Cal History
recalls each successful calibration. These dates may be reviewed by
scrolling with the UP / DOWN keys.
Input Type
indicates what kind of input or sensor the GM ECX is configured to accept and is
typically pre-configured at the factory. There are three Input Type possibilities consisting of
simple
EC negative,
simple
EC positive and Smart. Smart sensors upload sensor type and other
data to the GM ECX and may be viewed on the SMART SENSOR information screen.
New Sensor
displays the date when a new sensor was last installed.
Recent Cal
displays the most recent calibration date.
6.6
Clock / Warm-Up & Cal Purge Delays SETUP:
Since the GM ECX is equipped with a Real Time Clock & Calendar
Time
and
Date
must be set to
correctly match its location. They are set at the factory in a 24 hour format but may require
adjustment to match the location’s time & date after shipment. Follow the procedure in
Configuration Using the Magnetic Wand
in section 6.2.
Warm Up
and
Cal Purge
time delays are also available to prevent unwanted alarm trips. Figure
6-6 shows the menu for these items.