Revision 1.1
GM ECx Instruction Manual
25
Coefficient
Electrochemical Sensor Type
Default EUNITS
Negative
Negative
Negative
Negative, Bias
Negative, Bias
Negative
Negative
Negative
Negative
Negative
Negative
Negative
Negative
Negative
Negative
Positive
Positive
Positive
Hydrogen Sulfide
Oxygen
Carbon Monoxide
Ammonia
Nitric Oxide
Ethylene Oxide
Hydrogen Chloride
Hydrazine
Arsine
Sulfur Dioxide
Hydrogen
Hydrogen Cyanide
Phosgene
Phosphine
Hydrogen Fluoride
Nitrogen Dioxide
Ozone
Chlorine
ppm H2S
% Oxygen
ppm CO
ppm NH3
ppm NO
ppm Eth O2
ppm HCL
ppm N2H4
ppm Arsine
ppm SO2
ppm H2
ppm HCN
ppm COCl2
ppm PH3
ppm HF
ppm NO2
ppm Ozone
ppm Cl2
7.5
Zero Cal Value (Technicians only!):
The
Zero Cal Value
menu entry allows the zero calibration value to be set for something other
than a zero reading. In rare cases it may be necessary to perform Zero calibrations at some
other engineering unit reading than zero. Do not exceed 25% of full scale.
7.6
Raw Min / Max Counts (Technicians only!):
The
Raw Min / Max Counts
menus determine the range of 10-bit analog to digital (A/D)
converter counts that define 0 & 100% of full scale. Raw Min A/D counts create 0% readings and
Raw Max A/D counts create 100% readings. The corresponding
Zero 0%
and
Span 100%
readings are entered in the Configuration Menu describe in section 6.3. Live A/D count values
may be viewed from the CAL MODE Information screen described section 5.1.
REVISION HISTORY (after 3-13-2007)
Revision
1.0
Description
Initial Release
Date
7-1-07
Approved by:
DS
1.1
Updated per changes to wireless radio protocol
1-30-09
DS