GASMAX ECx Instruction Manual
Revision 1.1
18
again to load the new item and remove the cursor. Press NEXT to reverse out of the sub-menu.
To view
EC INPUT PAGE 2
, press the DOWN key with the pointer aimed at the bottom item on
PAGE 1.
Figure 6-2: Setup Menu Entry
6.3 System
Configuration
Menus:
The System Config. group consists of two pages of menus as shown in Figure 6-3. Each menu item’s
description follows in this section.
Figure 6-3: System Config. Menus
User Info
may be edited to contain virtually any 16-character ASCII field. It is typically used to
describe the monitored point by user tag # or other familiar terminology.
Eunits
(engineering units) may have up to a 10 character ASCII field. Many common gases have
pre-configured Eunits based upon the sensor type and each may be edited in this menu as
described in
Configuration Using the Magnetic Wand
section 5-2.
Zero (0%)
defines the reading to be displayed when 4mA (0%) is the GM ECX output.
Span (100%)
defines the reading to be displayed when 20mA (100%) is the GM ECX output.
The highest reading allowed is 9999 includes negative polarity sign and one decimal point.
Polarity is only indicated for negative readings.
Decimal Points
sets the resolution of the LCD readings and may be for 0, 1 or 2. Example:
ZERO readings for 0, 1 & 2 DP’s respectively are
0
,
0.0
&
0.00
.
Cal Span Value
sets what upscale value must be applied when performing Span calibrations.
Readout Deadband
allows forcing low values to continue to read zero. This is useful when there
are small amounts of background gases that cause fluctuating readouts above zero. The highest
amount of deadband allowed is 5%.
Track Negative,
set to NO, causes negative values to read the
Zero (0%)
value in data displays.
The CAL MODE readout displays negative values regardless of this setting and negative values
below the Fault set point will still cause the Fault alarm to trip. The gas value’s 200-1000 counts
output always locks at 200 counts when the reading is negative.
Backup Config.
allows users to store the entire current GM ECX menu database into non-
volatile memory for restoration later if incorrect values are accidentally entered or uploaded.