The application device for scanning is an optic-electronic device that captures high-resolution video images in real time during a patient
scan. It is connected to the laptop via a USB 2.0 high-speed cable for data transfer. The application device comprises of optics and sensor
electronics, which detect the real-time images which can then be further processed by the computer.
Figure 1: handpiece
Caution
Please do not look directly at the light source [because there is a possibility of causing a drop in visual acuity]
CRADLE
The cradle for the application device for scanning is an open case for the placement of the hand piece in resting position. I t contains a
heating plate responsible for the maintenance of the temperature of the glass plate at the head of the hand piece in order to avoid fogging
when used in the mouth. It is connected to PC with the USB cable.
Figure 2: cradle
Before mounting the hand piece into the cradle inlay part, the cradle inlay needs to be mounted into the cradle. When mounting the
handpiece into the inlay, two stabilization brackets will support the holding mechanism and the backside of the inlay.
Figure 3: USB cable
Figure 4: cradle inlay
Please note
The handpiece may not be actively scanning when placed into the cradle.
COMPUTER:
The high-performance computer has a graphics card on which the parallel graphics processes run in real time. The operating system and
user program must not be modified by the user. Software updates are managed by GC Europe N.V.