02_340_001_AJ_BE(Rev 11/05/11)
(from French version : 02_340_000_AJ_BE (Rev 11/05/11)
Portable Inspectra Laser
©Copyright Gazomat
29/48
7.5.4 S
ELECTING THE LANGUAGE
The user can select the device language from the Configuration menu.
Choice
Menu path
French
Configuration > Language > Français
English
Configuration > Language > English
German
Configuration > Language > Deutsch
Italian
Configuration > Language > Italiano
Dutch
Configuration > Language > Nederland
Hungarian
Configuration > Language > Magyar
Selecting the language
7.5.5 S
ELECTING THE MEASUREMENT MODE
The device has two measurement modes - absolute and relative - that can be selected from the
Measurement menu.
In Absolute mode
, the displayed concentration is the measured concentration as it is.
Absolute mode measurement
In Relative mode
, the displayed concentration is the measured concentration less the
baseline value. This mode is used to detect concentration variations in relation to an
ambient atmosphere containing methane in a slowly-changing concentration.
This mode is disabled when the measured concentration is greater than 50 ppm.
Relative mode measurement