02_340_001_AJ_BE(Rev 11/05/11)
(from French version : 02_340_000_AJ_BE (Rev 11/05/11)
Portable Inspectra Laser
©Copyright Gazomat
24/48
4
Using the pressure gauge of the flow
regulator, check that the residual
pressure in the canister is sufficient for
the check
5
Connect the gas check kit to the
detection device with the quick
coupler located at the end of the
connecting pipe. Insert the end piece
into the connector with a red ring on
the device.
6
Wait for a few seconds and then check
that the measurement displayed on the
device is within the tolerance range of
10 ppm +/- 2 ppm.
7
If the measurement is correct, the
device is in working order and ready
for use. The check is complete.
8
Disconnect the connecting pipe from
the detection device by pressing the
red ring while pulling the end piece of
the connecting pipe.
9
Take care to unscrew the flow
regulator from the gas canister to
prevent gas leaks
10
Important! Checking with the kit does
not replace annual calibration or
annual maintenance of detectors as
recommended by Gazomat.