5.0
PROBE
INFORMATION
Disposable Probes
•
DP400
Disposable Rectal/Esophageal -
Adult/Small Child (3' [0.9 meters] long, requires adaptor);
YSI400 series type
Reusable Probes
•
PAT101
Patient probe—Rectal/Esophageal -
Adult (10' [3.0 meters] long); YSI400 series type
•
PAT102
Patient probe—Rectal/Esophageal -
Pediatric (10' [3.0 meters] long); YSI400 series type
•
PAT108
Patient probe—Skin surface (10' [3.0 meters] long);
YSI400 series type
Probe Adaptor
•
ADP10
Reusable adaptor cable for DP400: connects Gaymar
disposable probe to Gaymar or Cincinnati Sub-Zero
control unit for all applications that call for Baxter/Pharmaseal
No. 66N2700.
•
ADP10B
Reusable adaptor cable for DP400: connects Gaymar
disposable probe to American Medical Systems control unit or
all applications that call for Baxter/Pharmaseal
No. 66N27100.
PROBE INFORMATION
WARNING
Some manufacturer’s patient probes may contain compensation resistors
in series with YSI400 series thermistors. Do not use these probes with
the Medi-Therm II machine.
Inaccurate patient temperature readouts will result and
inappropriate therapy may be delivered.
NOTE: Use YSI400 series patient probes or equivalent.
(Refer to the list of recommended probes above.)
7