w
w
w
.g
aymar
.com P/N 10187-000 7/98
To assure the optimum performance, dependability and safety, the following
should be performed every three (3) months or as specified in the facility's
preventive maintenance program.
An Inspection Form is provided at the end of this section to facilitate and
document the inspection process. Lower case letters preceding the subheadings
within section 7.3 correspond to the lines on the
INSPECTION FORM.
Always perform the
FUNCTIONAL CHECK AND SAFETY INSPECTION
after
making repairs and before returning the Medi-Therm II machine to
patient use.
Improper repair may result in death or serious injury,
equipment damage, or malfunction.
The following test equipment (or equivalent) is required to perform the
preventive maintenance/functional check procedures:
•
GAYMAR TPT9 Flowmeter/Temperature Tester
•
GAYMAR TFC1 Mercury Thermometer (±1°C accuracy);
-2°C to +52°C range
•
GAYMAR DBK9 Blanket Connector Hose
*
•
GAYMAR MT590 Medi-Therm II Test Tool, or
GAYMAR PRK2 Patient Temp Simulator Kit, or
Precision Decade Box (0-10K ohms, 0.2% accuracy, 1 ohm
increments)
•
GAYMAR DHP901 or DHP813 Hyper/Hypothermia Blanket
*
•
GAYMAR MT590 Medi-Therm II Test Tool, or
6" (approximately 150 cm) Shorting Jumper
•
Ground Resistance Checker
•
Current Leakage Tester
•
Static Control Wrist Strap
*
The GAYMAR MT590 Medi-Therm II Test Tool is a dedicated test tool
available from Gaymar Industries. This tool was designed to allow for more
convenient and safer testing of the high temp backup thermostats by
eliminating the need to remove the lower wraparound cover and avoiding
the use of a shorting jumper that could be accidentally left inside the
machine. In addition, the test tool can be used to monitor input/output
signals between the head and base of the machine through the use of
indicator lights. This feature can be of great value during any troubleshooting
process. This test tool also provides a patient probe simulator which allows
testing of the machine without a precision decade box.
For more information on the MT590 Test Tool, contact the Gaymar
Technical Service Department.
7.3 FUNCTIONAL CHECK &
SAFETY INSPECTION
FUNCTIONAL CHECK &
SAFETY INSPECTION
WARNING
Test Equipment
Requirements
20
SERVICE MANUAL
Medi-Therm II