SERVICE MANUAL
Medi-Therm II
REPAIR PROCEDURES
w
w
w
.g
aymar
.com P/N 10187-000 7/98
Figure 7—Flow Switch
CAUTION
Do not bend or alter terminals.
Terminals may break. Also, bending may alter the
trip point range.
9.
Mount thermostat, using existing screws.
10. Reapply insulating material.
11. Perform the
FUNCTIONAL CHECK AND SAFETY INSPECTION
(section 7.3).
To clean the flow switch (see figure 7 below and figure 15, pp. 74-75):
1.
Remove nuts (item A) and remove assembly.
2.
Loosen hose clamps (items B and C) and remove hoses.
3.
Remove barbed adapters (item D). Remove plunger.
4.
Clean all parts with alcohol and remove any foreign matter.
5.
To reassemble, reverse steps 1 through 3 above.
9.6
CLEANING THE
FLOW SWITCH
63