.com P/N 10187-000 7/98
Figure 6D—RFU Codes 4, 5
TROUBLESHOOTING CHARTS
41
RFU CODE 5
Compensation resistor 1
(R12 on
control/display board)
reading is too low
No
Yes
Repair/replace
the
control/display board.
Suspect
temperature
measurement
circuitry.
Is it
816 ohms
±0.1%
?
Replace R12.
Measure R12
resistance.
RFU CODE 4
Compensation resistor 1
(R12 on
control/display board)
reading is too high
SERVICE MANUAL
Medi-Therm II