background image

MGC Pump User’s Manual

UM-MGC-P v1.06 

7 of 19

DETECTOR RECORDS (LOGS) 

During operation, the detector records all usage activity. These records can be downloaded from the detector via 
the IR Link software and an IR Link or from the MGC Pump Dock.

The Event, Bump and Calibration Logs are always downloaded. The user may also choose to download different 
amounts of the data log to reduce the transfer time. Partial data logs contain approximately 1 weeks’ worth 
of data. New data log downloads will only contain the data since the last download. Full log downloads will 
download the entire data log, typically at least 2 months’ worth of data.

Event Log

The detector stores the last 25 alarm events. These are organized by first in, first out (FIFO) so the 26th event will 
replace the first event, and so on. The detector records the alarm conditions for each event:

•  Date and time at start of event
•  Duration of the alarm condition
•  Each sensor’s peak alarm status and reading

Bump Log

The detector stores the last 25 alarm events. These are organized by first in, first out (FIFO) so the 26th event will 
replace the first event and so on. Bump tests are differentiated from normal events when the alarm condition 
occurs inside of the MGC Pump Dock, or when the detector is manually bump tested according to the Manual 
Bump Test Instructions. The detector records the bump status for each test:

•  Date and time of bump test
•  Whether the test was performed manually or in an MGC Pump Dock
•  Each sensor’s peak alarm status and reading
•  The result of each sensor’s bump test
•  If the detector is displaying "BUMP DUE", the detector is either due for a bump test or has failed its last bump 
test

Calibration Log

The detector records the last 25 calibration attempts. These are organized by first in, first out (FIFO) so the 26th 
calibration will replace the first calibration and so on. Each calibration attempt will be recorded:

•  Date and time of calibration
•  Each sensor’s gas calibration concentration
•  Each sensor’s calibration success status
•  If the detector is displaying "CAL DUE", the detector is either due for a calibration because of a scheduled 
interval

Data Logs

The detector records its current operational status every second. The logging interval cannot be adjusted, but the 
detector compresses the data to reduce the storage and transfer times of redundant records. The typical logging 
capacity is at least 2 months of data. The following items are recorded into the log:

•  Date and time
•  Sensor readings and status conditions
•  All user and sensor options
•  Events (i.e. turn on or off)

Summary of Contents for Infrared

Page 1: ...User s Manual G A S C L I P T E C H C O M Multi Gas Clip Pump Infrared Multi Gas Clip Pump Pellistor...

Page 2: ...Detector Records Logs 7 Event Log 7 Bump Log 7 Calibration Log 7 Data Logs 7 How to Retrieve Logs Using the GCT IR Link 8 Detector Options Menus 9 Adjustible Options 9 Main Menu 11 Status Menu 11 Opt...

Page 3: ...nes lead compounds e g tetraethyl lead sulfur compounds phosphorus halogenated hydrocarbons and aerosols Exposure to poisons may impair the accuracy and or response time of the detector This applies t...

Page 4: ...imiter des nettoyants base d alcool et d agrumes des silicones des compos s de plomb par exemple plomb t tra thyle compos s soufr s phosphore hydrocarbures halog n s et a rosols L exposition des poiso...

Page 5: ...ion 1 Power Button 2 Audible Port 3 Display 4 Alarm Bar LEDs 5 Maintenance LEDs 6 IR Communication Window 7 Gas Inlet 8 Filter Window 9 Certification Label 10 Model and Serial Number 11 Gas Exhaust 12...

Page 6: ...bump test the calibration bottle will be displayed when it is time to apply gas The battery is displayed in 3 bars as well as a percentage The percentage calculation is approximate and can be used to...

Page 7: ...st LED flash Pump Blocked Fast beep Fast vibration Fast LED flash Sensor Error Fast beep Fast vibration Fast LED flash Low Battery 20 minutes remaining single beep flash and battery icon on 10 minutes...

Page 8: ...w for activation the detector will display LOW BAT and then turn off again Once activated the detector will begin to display gas readings immediately Each sensor will show a chasing O for the sensor r...

Page 9: ...the alarm condition occurs inside of the MGC Pump Dock or when the detector is manually bump tested according to the Manual Bump Test Instructions The detector records the bump status for each test Da...

Page 10: ...r 4 Select what kind of logs you want to download a Event Logs Only b Events Partial Data Log last week of data c Events New Data Logs Only All data logs from last time you pulled the logs to current...

Page 11: ...The detector s firmware version The TWA alarm limits The STEL alarm limits The LOW alarm limits The HIGH alarm limits Adjustable Options The detector s options are configured using the IR Link and th...

Page 12: ...ble Individual sensors can be disabled A disabled sensor is completely removed from the detector s display for sensor readings alarm limits and calibrations Caution A disabled sensor will not measure...

Page 13: ...empty the detector will skip ahead If the startup message does not fit on one screen it will scroll right to left twice Sub Menu Prompts Pressing the button during a sub menu prompt will cause the de...

Page 14: ...10 minutes remaining and then continuously for the last 5 minutes of run time When the battery has expired the detector will give 5 long beeps and flashes display OFF and then shut down To charge the...

Page 15: ...e detector will return to normal operation in alarm and the bump due date will be reset If a bump test interval has not been programmed into the detector users may simply apply gas to the detector dur...

Page 16: ...ternal pump and sensor areas free of water and debris If these filters become blocked then the detector will be unable to continue sampling and the pump will enter into a pump blocked alarm Looking at...

Page 17: ...e back of the detectors that can clip on to clothing or belts 6 Telescoping Probe P N MGC TSPROBE Metal probe that extends to 6 foot MGC Pump Probe Case P N MGC PUMP TSCASE Holster for MGC Pump with s...

Page 18: ...nal pump motor Replacement Particulate Filters P N MGC PFILTER 10 and MGC PFILTER 50 replacement particulate hard plastic filters in either a 10 pack or a 50 pack Replacement Hydrophobic Filters P N M...

Page 19: ...inuous 1 second data logging 2 months typical capacity Tests Full function self test upon activation Sensors battery and circuitry tests run continuously Ingress Protection IP 67 Warranty Full 2 years...

Page 20: ...ntrins que Class I Division 1 Groups A B C and D T4 Class I Zone 1 AEx d ia IIC T4 Gb 20 Ta 50 C 3 8Vdc nominal ANSI ISA 12 13 01 2000 C22 2 No 152 M1984 UL 916 4th Edition UL 913 8th Edition UL 60079...

Page 21: ...or use 3 Any damage or defects attributable to repair of the product by any person other than the authorized dealer or installation of unapproved parts on the product The obligations set forth in this...

Reviews: