MGC Pump User’s Manual
UM-MGC-P v1.06
1 of 19
WARNING STATEMENTS
Do not substitute components as this may interfere with the intrinsic safety of the device.
DO NOT substitute any other battery type than specified and supplied by Gas Clip Technologies.
Only use Gas Clip Technologies chargers and parts in the detector. Unapproved parts will void the warranty and
are considered unsafe.
Before each use check that all sensor and alarm ports are clear of any obstructions i.e. debris or blockage.
The detector contains a lithium battery that must be disposed of by a qualified recycler. Check local regulations
for proper disposal.
DO NOT charge the instrument in a hazardous location.
DO NOT use IR communications when an explosive atmosphere may be present.
If you suspect any malfunction or have any technical problems, contact GCT at 877-525-0808.
The battery may present a fire or chemical burn hazard if mistreated. Do not disassemble, heat above 100° C
(212°F) or incinerate. Contact Gas Clip Technologies for replacement instructions. Use of another battery may
present a risk of fire or explosion.
DO NOT charge the instrument in temperatures above or below the specified range of 0°C to 40°C
Keep new and used batteries away from children
DO NOT expose the detector to sensor poisons such as, but not limited to: alcohol, citrus-based cleaners,
silicones, lead compounds (e.g. tetraethyl lead), sulfur compounds, phosphorus, halogenated hydrocarbons
and aerosols. Exposure to poisons may impair the accuracy and/or response time of the detector. This applies
to pellistor version only.
Caution
: Before each day’s usage, sensitivity must be tested on a known concentration of methane equivalent
to 25-50% of full scale concentration. Accuracy must be within 0-20% of actual methane concentration.
Accuracy may be corrected by calibrating the detector.
If suspected sensor poisoning has occurred, recheck the detector (both calibrate and bump test).
The detector should be bump tested before use with a known concentration of gas to confirm its ability to
respond to gas. Calibrate the detector if the readings are not within the specified limits.
Any rapid upscale reading followed by a declining or erratic reading may indicate a gas concentration beyond
upper scale limit which may be hazardous.
Strong Electromagnetic Interference (EMI) may cause incorrect operations.
Only the combustible gas detection portion of this instrument has been assessed for performance by CSA
International.
READ FIRST BEFORE OPERATION
Gas Clip Technologies (GCT) Multi Gas Clip Pump (MGC Pump): detectors are personal safety devices designed
to detect the presence of specific toxic gases: carbon monoxide (CO), hydrogen sulfide (H
2
S), oxygen (O
2
) and
combustible gases/lower explosive limit (LEL). Before operation, ensure you have been properly trained on the use of
the equipment and appropriate actions to take in the event of an alarm condition.