Air filtration (B 1): 

The performance of the vacuum pump will reduce if the air filter is not maintained correctly.

 

 

The filter cartridge (8) should be cleaner by blowing air through the inlet to the outside of the cartridge, or if necessary 
replaced. A replacement of the filter is recommended at least once per year (or under extreme conditions more often). 
Filter cartridge change: Unscrew housing cover (7). Remove filter cartridge (8) and gaskets (9, 10) from filter chamber. 
Clean or replace filter cartridge (8) and check the condition of the gaskets (9, 10). Reassembly is in the opposite direction. 
Insert filter cartridge (8) with the closed end towards the housing cover (7). Pay attention that the gasket 9 is not confused 
with gasket 10 (which is thicker). Replace housing cover (7) and ensure the surface of both the rotor (13) and housing 
(12) are clean. Tighten the screws (17) to 5,5Nm +0,5Nm. 
 
 

Vanes (B 1 / B 2): 

The VTE pump has four graphite vanes (11), which have a low but permanent wear rate. The wear rate will be increased 
if the inlet air is contaminated. For pumping clean air the vanes should be check at the latest after 4000 running hours 
and thereafter every 1000 hours. For other conditions, depending upon the application, the service interval should be 
reduced. 
Checking the vanes: Remove the housing cover (7) from the housing. Remove the vane (11) to be checked. All vanes 
should have a minimum height (X) of 10mm (for VTE 3 and VTE 6) or 12mm (for VTE 8 and VTE 10). If the minimum 
height is reached then all four vanes should be replaced.  

 

The vanes should only be replaced as a complete set. 

Replacing the vanes: Clean housing (12) and rotor slots (13) with compressed air. 
Insert the vanes (11) in the rotor slots ensuring that the angled end of the vane (Y) is 
on the outside and in the direction of rotation (O1) along the bore of the housing (Z). 
Replace the housing cover (7) and lightly tighten. Start the pump briefly and check 
for the free and smooth running of the vanes. Tighten the housing cover screws (17) 
to 5,5Nm +0,5Nm.  

 

Fault Finding: 

Problem (P), Cause and Solution (S): 

P  Vacuum pump switches off due to activation of the motor protection switch.   
S  Voltage or frequency of supply does not correspond to the motor nameplate 

Æ

 ensure correct power supply 

S  Pump seized 

Æ

 Check if vanes are broken, otherwise contact local service centre 

S   Incorrect lead connection in terminal box 

Æ

 Check connections are according to diagram 

S   The back pressure of the motor fan is too high 

Æ

 Ensure sufficient cool ventilation 

S   The motor protection switch is not set correctly 

Æ

 Ensure settings are according the data onto motor nameplate 

S   Automatic cut-out activates too quickly 

Æ

 Use cut-out with time delay 

P  In sufficient vacuum flow: 
S  Inlet filter is blocked 

Æ

 Clean or replace the inlet filter 

S   Inlet or outlet pipe is too long or too restrictive 

Æ

 Reduce length or increase diameter of pipe

 

S  Leakage in pump or in system 

Æ

 Check / tighten fittings and pipes 

S  Worn or damaged vanes 

Æ

 Check vanes and replace if necessary 

P  Max. vacuum not reached: 
S  Leakage in pump or in system 

Æ

 Check / tighten fittings and pipes 

S  Worn or damaged vanes 

Æ

 Check vanes and replace if necessary 

P  Thermal switch activates 

Æ

 Vacuum pump is too hot: 

S  Temperature of the ambient air or inlet air too high 

Æ

 Reduced temperature of cooling or inlet air 

S  Pump is over loaded or set up incorrectly 

Æ

 contact local service centre 

S  Cooling air restricted 

Æ

 ensure sufficient cooling air 

P  Vacuum pump emits unusual noise: 
S  Pump housing is worn 

Æ

 Return pump to manufacturer or local service centre 

S  Vacuum regulation valve (if fitted) is noisy 

Æ

 Replace valve 

S  Vanes are worn or damaged 

Æ

 Check vanes and replace if necessary 

B 2

Summary of Contents for THOMAS PICOLINO VTE 3

Page 1: ...Betriebsanleitung Vakuumpumpe VTE 3 6 8 10 Seite 2 bis 5 Operation Manual Vacuum Pump VTE 3 6 8 10 Page 6 to 9...

Page 2: ...fenden Vakuumpumpen eignen sich zum F rdern von Luft mit einer relativen Feuchte von 30 bis 90 Es d rfen keine gef hrlichen Beimengungen z B brennbare oder explosive Gase oder D mpfe extrem feuchte Lu...

Page 3: ...er Temperatur von 20 C bis 60 C und einer Luftfeuchtigkeit von 30 80 zu lagern Bei einer relativen Feuchte von ber 80 empfehlen wir die Lagerung in geschlossener Umh llung mit eingelegtem Trockenmitte...

Page 4: ...t der schr gen Seite Y nach au en zeigen und diese Schr ge in Drehrichtung O1 mit dem Verlauf der Geh usebohrung Z bereinstimmt Geh usedeckel 7 leicht anschrauben Pumpe kurz einschalten und den freien...

Page 5: ...n Niederlassungen oder Vertragsfirmen in Anspruch zu nehmen Die Anschrift der f r Sie zust ndigen Service Stelle kann beim Hersteller erfragt werden siehe Hersteller Adresse Nach einer Reparatur bzw v...

Page 6: ...nning vacuum pumps are suitable for use with air of a relative humidity of 30 to 90 Dangerous gas mixtures e g inflammable or explosive gases or vapours extremely humid air water vapour aggressive gas...

Page 7: ...C Above 80 relative air humidity suitable additional packing with a drying agent should be used Initial operation B 1 The vacuum port 6 should be open or not restricted otherwise the vanes may break...

Page 8: ...es 11 in the rotor slots ensuring that the angled end of the vane Y is on the outside and in the direction of rotation O1 along the bore of the housing Z Replace the housing cover 7 and lightly tighte...

Page 9: ...to contact the manufacturer local subsidiary or representative for advice The address of the nearest service centre can be obtained from the manufacturer see manufacturer s address Following repair w...

Page 10: ...plug R ckschlagventil ZRK Non return valve ZRK Pos 6 Bild B1 Drawing B1 zwischen Pumpe und Schlauchanschluss between pump and pipe connector R ckschlagventil ZRK 12 Non return valve ZRK 12 R ckschlagv...

Reviews: