GANSHORN Medizin Electronic GmbH
Service Manual Ref.No. 03 140 0 046
PowerCube-Ergo-CASE_ENG_rev.02, 23.11.2012 Chapter 6, Troubleshooting, Checks & Adjustments, page 7 of 16
Flow Chart for Error An
alysis “Gas Calibration–Offset for O2/CO2 too High/ Low”
Check the actual calibration factors for O2- and/or CO2-Offset
Pict.: Calibration gas table, ergo calibration (gas calibration)
“Gas calibration offset”
Perform gas calibration correctly after warm-up time
of 30 mins. (see instruction manual) OK?
YES
Check the O2-Offset / CO2-Offset. OK?
NO
YES
1) Adjust the O2-Offset / CO2-Offset
2) Replace the O2-Analyser / CO2-Analyser
NO
OK