GANSHORN Medizin Electronic GmbH
Service Manual Ref.No. 03 140 0 046
PowerCube-Ergo-CASE_ENG_rev.02, 23.11.2012
chapter 3, Disassembly, page 7of 10
_____________________________________________________________________
How to Remove / Replace the Suction Pump
1) Remove the cover correctly from the base assembly
The suction pump tubes remain attached to the pump, do not remove!
2) Detach the suction pump tubes
a. from the compensation vessel (tube coming from pump
connector “V”) and
b. from the OUT-socket in the PowerCube
’s rear panel (tube coming from the pump
connector “P”)
3) Cut the cable tie holding the suction pump and unplug the pump cable from the Controller
module
4) Withdraw the suction pump
To assemble the modules again perform the same steps backwards
Please take care to ensure the correct flow direction of the suction pump when connecting the tubing:
a. To the OUT-
socket in the PowerCube’s rear panel connect the tube coming from the P-
marked side of the suction pump (P=pressure)
b. To the Compensation vessel connect the tube coming from the V-marked side of the
suction pump (V=vacuum)
After replacement of the suction pump please perform
1) Suction flow adjustment (see chapter 6, Troubleshooting, checks and adjustments)
Pict.: PowerCube-Ergo, inner lateral: with compensation vessel (1) and suction pump (2).
Other components: CO2-Analyser (3) and calibration valve (4)