GANSHORN Medizin Electronic GmbH
Service Manual Ref.No. 03 140 0 046
PowerCube-Ergo-CASE_ENG_rev.02, 23.11.2012
chapter 3, Disassembly, page 2of 10
_____________________________________________________________________
Disassembly
Warning notices
DANGER
REQUIREMENT FOR SERVICE PERSONEL
Opening the PowerCube-housing as well as all changes, modifications,
extensions, repairs, maintenance measures, and any other work on GANSHORN
products including meteorological / safety checks may only be performed by
GANSHORN or by an authorized GANSHORN partner.
Only qualified service technicians that have been trained on repair and
maintenance of the PowerCube are authorized to perform the procedures
described in this manual!
DANGER
DISCONNECT FROM THE MAINS BEFORE ANY REPAIR WORK
Disconnect the PowerCube from the mains before starting any repair work.
If checks and adjustments of the PowerCube-module can only be carried out by
removing the cover AND when connected to the mains, it will be pointed out
especially in the respective description.
When performing these procedures be aware that parts inside the device may be
current-bearing.
CAUTION
ELECTROSTATIC DISCHARGE (ESD)
The PowerCube contains electro-static sensitive electronic components;
observe antistatic precautions for any maintenance procedure:
The unit must be placed on an earthed antistatic mat
Service personnel must be earthed and have to wear anti-static
clothing/shoes when handling boards or components
Comply with the international standard IEC 61340-5-1
Always use an antistatic bag when transporting boards or components