GANSHORN Medizin Electronic GmbH
Service Manual Ref.No. 03 140 0 046
PowerCube-Ergo-CASE_ENG_rev.02, 23.11.2012
Chapter 1, safety regulations, page 4 of 8
DANGER
INSTALLATION, UNINSTALLING
All components of the GANSHORN device have to be switched off and
disconnected from the mains for assembling and installation, as well as for
uninstalling and disassembling.
Installation and initial operation and any other work on GANSHORN products
including meteorological / safety control may only be performed by GANSHORN
or by an authorized GANSHORN partner.
DANGER
MODIFICATIONS, MAINTENANCE AND SERVICE
If the GANSHORN device is modified, appropriate inspection and testing must
be conducted to ensure continued safe use of the device.
After each maintenance check or other operation with the device or the system a
metrological and safety check is necessary. Each metrological check demands a
succeeding safety check.
All changes, modifications, extensions, repairs, maintenance measures, and any
other work on GANSHORN devices including meteorological / safety checks may
only be performed by GANSHORN or by an authorized GANSHORN partner.
Already changing the power supply is mentioned as a modification.
CAUTION
DANGER OF INJURY - INSTALLATION
Wear protective gloves and safety shoes for installation and transport!
In order to ensure maximum protection during transport, all GANSHORN devices
are packed at the factory in suitable cartons and/or boxes with sufficient internal
packing. Please observe the graphical symbols and the instructions on the
packaging.
CAUTION
DATA SAFETY
In the case of a computer defect, the patient data, incl. measurement results,
can go lost irrevocably!
Backup archives regularly to an external storage media (ZIP-drive, streamer,
CD-ROM, recording system, etc.)!
=> Instruction manual Software LF8,
Chap. 4 Data Management!
WARNING
GAS MIXTURE CHECK
Before using the gas mixture, the values of the individual gas components have
to be checked with regard to the requirements of the measurement to be
performed
WARNING
PRESSURIZED GAS BOTTLES
Improper handling of gas bottles creates a potential danger for people and
objects! The relevant regulations are to be observed!
Secure the bottles, especially to avoid accidents!
The content and filling condition must be clearly recognizable at all times!
Gas bottle seals, connections and lines must be free of oil and grease!
Have the gas bottle regularly inspected by an official inspection agency!
The operator is liable! We recommend rented gas bottles.
Close the main cock whenever the gas bottle is not in use!