12.2

 

RS232 Pinout (9-pin D-Sub) 

 

DE-9 plug pin-out (9-pin., male), as it is usually 

available at DTE 

 

DE-9 socket pin-out (9-pin., female), as it is 

usually available at DCE 

DTE - data terminal equipment (Terminals resp. computer) 
DCE - data circuit-terminating equipment (Balance) 
 

Signal 

Pin 

Origin 

Abbreviation  Name 

Typical purpose 

DE-9  DTE (e.g. PC)  DCE (e.g. Balance) 

RxD 

Receive Data 

Carries data from DCE to DTE.  2 

Input 

Output 

TxD 

Transmit Data  Carries data from DTE to DCE.  3 

Output 

Input 

GND 

Ground 

 

 

12.3

 

Description of the data transfer  

Symbols and composition of the data 

-/+ 

2 bit 

Negative-sign, if the value is positive a space is transmitted. 

Data  7 bit 

Value for weight / decimal point 

Unit  3 bit 

Weight unit  

CR 

1 bit 

Carriage Return 

LF 

1 bit 

Line Feed 

 
Example:  
The output-value: -12,345 kg 
Output data, displayed as ASCII code 

-/+ 

Data 

UNIT 

CR 

LF 

 

 

 

 

\r 

\n 

 

12.4

 

Remote control instructions 

Via the RS232 data interface, the computer can control the balance with subsequent commands.  
If the transmit / device setting is 27 (27=1B

hex

, default setting), the remote control commands are sent to 

the balance by HEX-code: 
 
1B 70 (ACSII Code p): Send data 

 

 

(Print)  

1B 71 (ACSII Code q): calibration function is active  (CAL - Key) 

 

1B 72 (ACSII Code r): Counting function is active 
1B 73 (ACSII Code s): Weighing unit selection 
1B 74 (ACSII Code t): Taring   

 

 

(TARE - Key) 

 

Drivers and software can be free downloaded on our homepage. 

www.gandg.de

 

Summary of Contents for E100YB-C

Page 1: ...6 0 7 Version 2 0 04 2015 EN Precision balances EY E150Y E300Y E600Y E1200Y E3000Y E6000Y E150Y C E300Y C E600Y C E1200Y C E100YB C E200YB C E300YA E600YA E1200YA E3000YA E6000YA E2000YB E3000YB To e...

Page 2: ...acking Installation 9 6 3 General power supply 10 6 3 1 Battery power supply 10 6 3 2 Mains connection during rechargeable battery operation 10 6 3 3 Battery exchange 10 6 4 Initial Commissioning 10 6...

Page 3: ...12 1 Technical Data 17 12 2 RS232 Pinout 9 pin D Sub 18 12 3 Description of the data transfer 18 12 4 Remote control instructions 18 13 Maintenance and Servicing 19 13 1 Cleaning 19 13 2 Maintenance...

Page 4: ...1 d 1 d Warming up time 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz Unze GN Gran dwt Pennyweight l...

Page 5: ...ming up time 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz Unze GN Gran dwt Pennyweight lb Pfund LCD...

Page 6: ...ght 1 d 1 d 1 d 1 d 1 d 1 d 1 d Warming up time 30 min 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz...

Page 7: ...or battery accumulator 2 1 Proper Use The balance you purchased is intended to determine the weighing value of material to be weighed It is intended to be used as a non automatic balance i e the mater...

Page 8: ...afety Precautions COUTION Carefully read this operation manual before setup and commissioning Environmental conditions only use indoor max Altitude 2000m Operating temperature 20 C 20 C humidity of ai...

Page 9: ...ou select the right location for your balance Therefore observe the following for the installation site Major display deviations incorrect weighing results may be experienced should electro magnetic f...

Page 10: ...dapter Otherwise there is a risk that the Balance be impaired If the balance is not used for a long time remove the battery accumulator and store them separately Leaking battery liquid could damage th...

Page 11: ...technical measuring Info about adjustment weights can be found on the Internet at www gandg de Operations sequence Observe stable environmental conditions A warming up time 30 min is required for stab...

Page 12: ...Note The PVC wind protection is supplied for model E150Y E300Y E600Y E1200Y E300YA E600YA und E1200YA see chap 1 Technical data Note The Glass wind protection is supplied for model E150Y C E300Y C E6...

Page 13: ...G G GmbH Operating elements 13 8 Operating elements Pic 1 E150Y E300Y E600Y E1200Y E300YA E600YA and E1200YA Pic 2 E3000Y E6000Y E3000YA E6000YA E2000YB and E3000YB...

Page 14: ...lance before the first commissioning Level the balance with the adjustable feet horizontal until the air bubble in the level is in the center of the Level Indicator Give the balance after changing the...

Page 15: ...parts it must know the average part weight i e reference Proceed by putting on a certain number of the parts to be counted The balance determines the total weight and divides it by the number of part...

Page 16: ...re changes must be made only at special conditions of use Overview In the off state press the CAL button and keep it pressed press the ON OFF button turn on the balance Now the balance is accessed to...

Page 17: ...enable Default setting is 1 h Save settings After all points have been configured press CAL button returns the balance automatically into weighing mode i Restore the factory settings In the off state...

Page 18: ...if the value is positive a space is transmitted Data 7 bit Value for weight decimal point Unit 3 bit Weight unit CR 1 bit Carriage Return LF 1 bit Line Feed Example The output value 12 345 kg Output...

Page 19: ...ual vacuum cleaner Remove spillages immediately 13 2 Maintenance and repair The appliance may only be opened by trained service technicians who are authorized by G G Before opening disconnect from pow...

Page 20: ...not at zero Adjustment is no longer correct Great temperature fluctuations Warm up time was ignored Electromagnetic fields static charging choose different location switch off interfering device if p...