G&G GmbH | Basic Operation
14
Pic.3 E150Y-C, E300Y-C, E600Y-C, E1200Y-C, E100YB-C and E200YB-C
1. Weighing pan*
2. Level Indicator
3. ON/OFF-key
4. Calibration-key
5. Unit-key
6. Count-key
7. Light-key
8. Tare-key
9. Adjustable feet
10. LCD-Display
11. RS-232 Interface
12. Power socket
13. Pic.1 PVC wind protection
14. Stainless steel column
15. Wind protection ring
Pic.3 Glass wind protection
Note: The weighing pan is supplied depending on the model in a round or square version. (see
9
Basic Operation
After new purchase of the balance or after a long storage, you have to calibrate the balance before the first
commissioning. Level the balance with the adjustable feet horizontal until the air bubble in the level is in
the center of the Level Indicator. Give the balance after changing the location a warm-up time min. 30 min.
before using. This time is necessary for the balance to adapt to the eventual temperature.
9.1
ON/OFF - [Power On/OFF]
Press the [ON/OFF] button to turn on the balance. The display shows the message "
ON
". Then the max.
weighing range is displayed. The balance will carry out a self-test (
F----1
to
F----9
). When the "
0.00
g" in
display appears, the balance is ready for operation.
Press the [ON/OFF] button and keep it pressed, until the display shows the message “
OFF
”. The balance
will switch off then.
9.2
TARE
When the pan is empty and indication is different than zero press [TARE] key.