G&G GmbH | Displayinformation / Error messages 

16 

 

10

 

Displayinformation / Error messages 

F----1

   

Date over display scope 

F----2

   

A/D-switch disorder 

F----3  

 

Function key disorder 

F----5  

 

Sensor signal disorder / Load Cell defect  

F----L

   

Lower zero position of sensor signal 

F----H

   

Weighing capacity is exceeded 

C----F 

 

Calibration error, instability or not at zero point during calibrating 

C----H

  

 

Calibration error, items on the pan or at higher zero point during calibrating 

E----- 

 

Storage calibration data error, re-calibration is necessary 

------

   

Zero point of the scale disorder / mechanical defect of the scale 

 

 

Should other error messages occur, switch balance off and then on again. If the error 
message remains inform manufacturer. 

11

 

Configuration & User Menu 

The settings of the balance can be changed in the setting menu. This way, the balance can be adjusted to 
individual weighing requirements. By the factory the user menu has been set in a way that normally no 
more changes must be made, only at special conditions of use. 
 
Overview: 
In the off state press the [CAL] button and keep it pressed, press the [ON/OFF] button turn on the balance. 
Now the balance is accessed to the setting menu. Using the [CAL] button the individual menu items (

C1-

C3

) can be selected one after the other. Press the [TARE] button to change the setting value. After all 

points have been configured, press [CAL] button returns the balance automatically into weighing mode. 
 
a)

 

C1 = sensitivity setting - values: 0 1 2 3 4  
The higher the value, the lower the sensitivity and higher stability.  
Filter 0: Setting for dispensing. 
Filter 1/2: Sensitive and fast, very quiet using location. 
Filter 3/4: Robust but slow, busy using location.  
Default setting is 2 or 1.  

 
b)

 

C2 = filtering strength setting - values: 0 1 2 3  
The higher the value, the lower the sensitivity and higher stability. 
Filter 0: Setting for dispensing. 
Filter 1/2: Sensitive and fast, very quiet using location. 
Filter 3: Robust but slow, busy using location.  
Default setting is 2 or 1.  

 

c)

 

C3 = serial port setting - transmission speed (RS-232) 
2(600), 3(1200), 4(2400), 5(4800), 6(9600) - Default setting is 6.  

 

Summary of Contents for E100YB-C

Page 1: ...6 0 7 Version 2 0 04 2015 EN Precision balances EY E150Y E300Y E600Y E1200Y E3000Y E6000Y E150Y C E300Y C E600Y C E1200Y C E100YB C E200YB C E300YA E600YA E1200YA E3000YA E6000YA E2000YB E3000YB To e...

Page 2: ...acking Installation 9 6 3 General power supply 10 6 3 1 Battery power supply 10 6 3 2 Mains connection during rechargeable battery operation 10 6 3 3 Battery exchange 10 6 4 Initial Commissioning 10 6...

Page 3: ...12 1 Technical Data 17 12 2 RS232 Pinout 9 pin D Sub 18 12 3 Description of the data transfer 18 12 4 Remote control instructions 18 13 Maintenance and Servicing 19 13 1 Cleaning 19 13 2 Maintenance...

Page 4: ...1 d 1 d Warming up time 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz Unze GN Gran dwt Pennyweight l...

Page 5: ...ming up time 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz Unze GN Gran dwt Pennyweight lb Pfund LCD...

Page 6: ...ght 1 d 1 d 1 d 1 d 1 d 1 d 1 d Warming up time 30 min 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz...

Page 7: ...or battery accumulator 2 1 Proper Use The balance you purchased is intended to determine the weighing value of material to be weighed It is intended to be used as a non automatic balance i e the mater...

Page 8: ...afety Precautions COUTION Carefully read this operation manual before setup and commissioning Environmental conditions only use indoor max Altitude 2000m Operating temperature 20 C 20 C humidity of ai...

Page 9: ...ou select the right location for your balance Therefore observe the following for the installation site Major display deviations incorrect weighing results may be experienced should electro magnetic f...

Page 10: ...dapter Otherwise there is a risk that the Balance be impaired If the balance is not used for a long time remove the battery accumulator and store them separately Leaking battery liquid could damage th...

Page 11: ...technical measuring Info about adjustment weights can be found on the Internet at www gandg de Operations sequence Observe stable environmental conditions A warming up time 30 min is required for stab...

Page 12: ...Note The PVC wind protection is supplied for model E150Y E300Y E600Y E1200Y E300YA E600YA und E1200YA see chap 1 Technical data Note The Glass wind protection is supplied for model E150Y C E300Y C E6...

Page 13: ...G G GmbH Operating elements 13 8 Operating elements Pic 1 E150Y E300Y E600Y E1200Y E300YA E600YA and E1200YA Pic 2 E3000Y E6000Y E3000YA E6000YA E2000YB and E3000YB...

Page 14: ...lance before the first commissioning Level the balance with the adjustable feet horizontal until the air bubble in the level is in the center of the Level Indicator Give the balance after changing the...

Page 15: ...parts it must know the average part weight i e reference Proceed by putting on a certain number of the parts to be counted The balance determines the total weight and divides it by the number of part...

Page 16: ...re changes must be made only at special conditions of use Overview In the off state press the CAL button and keep it pressed press the ON OFF button turn on the balance Now the balance is accessed to...

Page 17: ...enable Default setting is 1 h Save settings After all points have been configured press CAL button returns the balance automatically into weighing mode i Restore the factory settings In the off state...

Page 18: ...if the value is positive a space is transmitted Data 7 bit Value for weight decimal point Unit 3 bit Weight unit CR 1 bit Carriage Return LF 1 bit Line Feed Example The output value 12 345 kg Output...

Page 19: ...ual vacuum cleaner Remove spillages immediately 13 2 Maintenance and repair The appliance may only be opened by trained service technicians who are authorized by G G Before opening disconnect from pow...

Page 20: ...not at zero Adjustment is no longer correct Great temperature fluctuations Warm up time was ignored Electromagnetic fields static charging choose different location switch off interfering device if p...

Reviews: