3 - 8
Wait until some curvature is seen in the temperature versus time curve. Figure 3-1 shows a good
quality curve. Select F2-Skip to terminate the experiment.
g)
Turn off the heater by setting the power to 0%.
h)
Close the CPT110 Runner Window by pressing F2-Skip.
4. Graphing the Temperature versus Time Curve
a)
Run Microsoft Excel.
b)
Select
File, Open...
from the Excel menu. In the resulting dialog box, select the "TUNE.DTA" file
in your Gamry Framework data directory. Select
Ok
.
NOTE: The data curve resulting from the TDCTUNE experiment is normally stored in the file
"TUNE.DTA". If you stored the data in a different file, open that file.
c)
The "TUNE.DAT" file is not in an Excel format, so a series of dialog boxes will pop up on the
screen. Each dialog box asks a question concerning the format of the file. Select
Next>
or
Finish
in each box to accept the default answer to each question.