Back Panel Operation
Description
DIGITEL MPCe Rear Panel
Input Power, J507
The J507 location is dedicated to input power. It is a standard IEC interface. Prior to attaching input power, verify
the proper voltage of the controller is 110 or 220-230 volts. This information is available on the serial number
block of the controller and is configured in the part number. Gamma Vacuum places a label in proximity to the
J507 input stating the wired input voltage upon shipment.
Contact Gamma Vacuum if a change in input power is required.
10kV SHV/SAFECONN, J401-J504 Pairings
A 10kV SHV and corresponding SMB connector is required for each high voltage connection on the rear panel of
the controller. 10kV connections are used to relay high voltage to the ion pump. The SMB connections are used
to relay the SAFECONN signal from the MPCe/LPCe to the ion pump high voltage feedthrough.
10kV SHV connector operates similarly to BNC style connectors. Insert the mating connector the desired J50X
connection and rotate clockwise until the connection is snug.
SMB connectors are push on connectors. Press firmly until the connector snaps into place.