MPCe/LPCe Users Manual
14
Identification/Calibration Information
The controller indicates identification and calibration information across the top of the screen. “Gamma Vacuum”
along with the date, time, and model are permanently displayed information. In addition, the controller can display
up to 3 symbols representing areas of the controller that remain to be calibrated. Those are:
1. The touch screen calibration indicator is symbolized by cross hairs
2. The high voltage calibration indicator is symbolized by the letters “HV”
3. The TSP calibration indicator is symbolized by the letters “TSP” on controllers with this option installed.
When the MPCe is linked to the remote controller, a line with arrows on each end is present. If the TSP is
not present or if the type of controller has not been selected in the TSP Config Menu, the line with arrows
will have a slash through it.
4. A HV Interlock icon is displayed if the HV1 or HV2 interlocks are not preset.
5. A TSP Interlock icon is displayed if the TSP interlock is not present.
6. The capital letter F, L, or R signify the communication modes of:
a. F = Full, both front panel and serial commands are accepted (default).
b. L = Local, only front panel commands are accepted.
c. R = Remote, only serial commands are accepted.
7. A pad-lock symbol is shown when the touch screen is locked.
Supply Sections
Current, pressure, and voltage are displayed simultaneously. The desired reading can be rotated into the main
position by pressing the value button below the Start/Stop HV button.
If a supply is not installed, the reading for that supply will indicated “NOT INSTALLED”. A second section can be
added in the factory or by a trained technician.
Menu Button
The menu button opens the main menu of the MPCe. This button does not exist on the LPCe. The LPCe Menu
is always visible.