MPCe/LPCe Users Manual
10
Required Items
You need the following items to install the controller:
1. 3-wire, detachable universal input power cable (included with controller)
2. high voltage (HV) cable for each pump (ordered separately)
3. safety ground cable for each pump (included with high voltage cable)
Procedure
Perform the following procedure to install the controller (see Back Panel Description for real panel image):
1. Place the controller in its location and secure as necessary.
2. Connect the safety ground cable between the pump and safety ground stud at the DIGITEL rear panel.
3. Connect the high voltage cable to the ion pump and to the appropriate high voltage connector on the DIGITEL
rear panel (J501-J504).
4. If you have the optional SAFE-CONN feature, an additional connector is part of the HV connector. Connect it
to the appropriate SAFE CONN connector (J401-J404).
5. Verify correct input voltage requirements. Connect the AC input power cable to input power receptacle on the
DIGITEL rear panel.