Revision 1
28
9/10/2019
Section 3
Installation
3.1 Receiving the System
When the system arrives, inspect the packaging for external signs of damage. If there is any
obvious physical damage, contact the shipping agent and Galvanic Applied Sciences to
report the damage and request that the carrier's agent be present when the unit is unpacked.
It is recommended that you retain the shipping container so that it may be used for future
shipment of the unit, if necessary.
3.2 Installation Requirements
3.2.1 Electrical Requirements
The analyzer can be powered by 100-240 VAC or 10-32 VDC sources. The power
requirement is 10 W.
If desired, it is possible to connect both AC and DC power supplies. In this manner, the DC
power supply as a backup power. When choosing a power source, the type of solenoid
drivers need to be considered (AC or DC).
3.2.2 Location of the System
The system is designed to be operated at ambient temperatures from 10 - 50
o
C. Galvanic
Applied Sciences offers complete analyzer shelters from sun shades to complete buildings,
please contact Galvanic Applied Sciences, Inc. (or your local representative) for additional
information.
The system should be mounted in a location where it is not exposed to excessive vibration.
In addition, it should be mounted in a location where the ambient pressure remains relatively
consistent over the course of a day
The vent line cannot have any backpressure on it, as backpressure can seriously impact
readings. The vent line should be as short as possible at a slight continuous downward slope.
3.2.3 Space Requirements
The size of the analyzer is 17” H x 19” W x 9” D (43 cm x 48 cm x 23 cm). Mounting
information is provided in Figure 3-1.
Summary of Contents for ProTech903W
Page 8: ...Revision 1 8 9 10 2019...
Page 109: ...Revision 1 109 9 10 2019 Figure 5 34 Create Modbus List Dialogue Box...
Page 112: ...Revision 1 112 9 10 2019 Figure 5 36 Typical Modcon with Floating Point List...
Page 141: ...Revision 1 141 9 10 2019 Figure 6 18 Status Tab...
Page 171: ...Revision 1 171 9 10 2019...
Page 172: ...Revision 1 172 9 10 2019 Section 13 MSDS Information...