Revision 1
93
9/10/2019
5.10.3.2 Switching Field
The
Switching
field (Figure 5-22) allows the user to configure what relays and or solenoids
are associated with each of the gas run types that are being used. There are 6 relays and 8
solenoids that can be controlled by the analyzer. Putting a checkmark in any of the relay or
solenoid boxes will cause that relay or solenoid to energize when one of the analysis modes
is running.
Figure 5-22: The Switching Field
5.10.3.3 Frequency
If the
Time
check box is checked (see Section 5.10.3.1) the operator can schedule the gas
run as per set in the Frequency section (Figure 5-23) of the gas run type tab. The drop down
Day of Week list includes Everyday and each day (Sunday to Saturday).
Figure 5-23: The Frequency Field
5.10.3.4 Analog Output Override
If an
Analog Output Override
if enabled (Figure 5-24), the user can set up two analogs
channels for each of the gas run types that will override the default analog outputs set on the
Outputs
Tab. An example of this would be to set the
Function
of analog output
Channel
1 to
Stream 1 Calibration
if the default was set up as “
Stream 1 Sample
. This would allow Analog
Output 1 to be used for both Stream 1 Sample Reading and Calibration Reading. The
Function dropdown list includes stream readings and a variety of other parameters and the
Channel list includes all six analog outputs.
Figure 5-24: Analog Output Field
Summary of Contents for ProTech903W
Page 8: ...Revision 1 8 9 10 2019...
Page 109: ...Revision 1 109 9 10 2019 Figure 5 34 Create Modbus List Dialogue Box...
Page 112: ...Revision 1 112 9 10 2019 Figure 5 36 Typical Modcon with Floating Point List...
Page 141: ...Revision 1 141 9 10 2019 Figure 6 18 Status Tab...
Page 171: ...Revision 1 171 9 10 2019...
Page 172: ...Revision 1 172 9 10 2019 Section 13 MSDS Information...