Revision 1
100
9/10/2019
5.12 Inputs - WET Tab
Changing parameters on this tab may affect the operation of the system.
Any changes should be performed by an authorized operator.
There are four isolated wet contact discrete inputs on the analyzer. The voltage input range
required for these inputs is 4 - 24 VDC. The
Inputs - WET
tab allows the user to configure
these discrete inputs in any manner. These inputs can be configured to trigger a certain type
of run, or to monitor external pressure or temperature switches, for example. The
Inputs- Wet
tab is shown in Figure 5-27. Each discrete input is configured in the same manner.
Figure 5-27: Inputs - Wet Tab
The configuration and section descriptions for this tab are identical to the
Inputs-DRY
tab
(Section 5.10).