Troubleshooting

 

 

 

If in doubt contact service support on +44 (0)1527 400022. 
• 

Seat Does not move 

Check handset 

– there should be a light as shown in the images 

 
 

To test the RCCB push the test button, the switch should trip. Push the RESET button; power 
should be restored. If the RCCB fails to operate in this manner there is a problem with the electric 
supply. When mains power is interrupted the bath will only work on battery backup to allow a user 
to exit the bath and will then stop. Lower the bath and do not use until the fault is rectified. 
 
Once power is restored allow the batteries to recharge before operation (usually 3 hours) 
If the bath fails to operate, reset the system: Depress both seat buttons on the handset at the 
same time 

– 10 long beeps will sound and the system will reset. It is worthwhile checking that the 

handset cord and jack are still firmly located into the control box. 
 

Bath judders: 

usually occurs if the bath has not been fixed to the floor correctly or the bath is not 

level. 

 
Bath noisy: 

check to ensure there is nothing touching the bath while it moves. Ensure that the 

pipe-work and cables are free of the panels and not rubbing against the framework. Check panels 
are free to move and are not touching each other. 
 

Bath leaking: 

look for source of leak-typically caught flexi-hose on feed or waste. This can usually 

be remedied by your own maintenance contractor. 
 

Bath beeping: 

mains power has failed to the unit- 

see section titled ‘Bath does not move’. The 

handset

  

‘battery’ symbol should be illuminated when ‘Seat in’ & ‘Seat out’ button pressed. 

 
 
 
 
 
 

Tel: +44 (0)1527 400022 

email: 

[email protected]

 

 

Summary of Contents for TALANO

Page 1: ...ser Manual v1 4 Gainsborough Bathrooms Gainsborough House Brickyard Road Aldridge WS9 8SR United Kingdom Tel 44 0 1527 400022 email service gainsboroughbaths com www gainsboroughbaths com TALANO ALERA...

Page 2: ...larly with a medical disinfectant as well as following a strict cleaning routine In the event of mains failure during use of the bath back up batteries will enable the bather to exit the bath The bath...

Page 3: ...rature of the hot water is limited to 43 C The water temperature can be reduced by adding cold water We suggest that while the bath is filling the seat is brought out of the bath and lowered ready for...

Page 4: ...eat out button Step 2 Place the bather onto the seat Fig 1 Ensure that the lap strap has been fitted and seat arms are placed into the lowered position Step 3 Lift the bather into the bath Fig 2 Fig 3...

Page 5: ...e chair Client Transfer from Seat Trolley to bath 1 Wheel the client to the bath on the transfer seat and trolley Ensure that the transfer arm is fully lowered using the Seat Out button The trolley sh...

Page 6: ...the seat fully using the Seat out button on the handset 3 When the seat is fully down and all movement has stopped gently push the upper safety latch down to release Fig 6 4 The transfer trolley is th...

Page 7: ...re pressing the button Temperature Adjustment The red fill button supplies hot water at a pre set temperature usually 43 C To reduce the temperature of the bath water during the autofill process pres...

Page 8: ...e At this stage the leg lift will still be in its lowered position Fig 1 When the seat has rotated approximately 45 the leg lift will begin to operate Fig 2 Once the legs have been fully lifted the se...

Page 9: ...olour The colour therapy lights offer a choice of settings as detailed Colour options are as follows Green Orange Red Violet Blue Yellow Indigo Rainbow Spectrum rotating lights Off Simply press to tur...

Page 10: ...the seat to the bath should be cleaned using a silicone based furniture polish i e Mr Sheen Please do not use wax based products Wipe clean and leave no residue after application Weekly Check water t...

Page 11: ...all over ensuring there are no exposed wires or pipes Panels should all be fitted tightly with screws and fixings in place Load the seat to its SWL Operate the bath through its full sequence The seat...

Page 12: ...eat buttons on the handset at the same time 10 long beeps will sound and the system will reset It is worthwhile checking that the handset cord and jack are still firmly located into the control box Ba...

Reviews: