background image

 

Section 10 

 

00-02-0511 

2019-07-10 

Setup (continued)_________________________________________________

Use the UP and DOWN buttons to select the desired channel to 
set up. (Select CH 0 to exit.) Press SCROLL ENTER, and the 
LCD will read: 

 

Use the UP and DOWN buttons to adjust the offset for the 
channel being set. Press the SCROLL ENTER button to go back 
to the channel display. Go to CH 0 to exit by pressing SCROLL 
ENTER. Press the UP button to go to SETUP 3. 

Setup 3 SETPOINTS 

Press the SCROLL ENTER button to get into the setpoint setup. 
The LCD will first display the Channel Select entrance/exit 
display: 

 

Use the UP and DOWN buttons to select the desired channel to 
setup. Press the SCROLL ENTER button and the LCD will read: 

 

SP# indicates the output that the setpoint is tied to. For example, 
SP 1 is tied to Output 1 (SP1 OUT). Use the UP and DOWN 
buttons to set the desired set point. Press the SCROLL ENTER 
button to accept the entry. Repeat for SP2 and SP3, and you will 
return to the Channel Select display. Repeat until all set points 
are entered. 

Note that entering Ø (zero) for a setpoint will disable the output for 
that particular channel. With zero as the set point, open 
thermocouple will not cause that output to activate. In this way, 
channels can be grouped, if desired. Further, an alarm before 
shutdown feature can be implemented by setting the set points 
progressively higher. For example, if SP1 = 900 and SP2 = 1000, 
Output 1 (SP1 OUT) could be wired to an Alarm Input and Output 
2 to a Shutdown Input on an annunciator. The range of 
adjustment is 0-1999 in the Units (°F or °C) chosen for those 
channels. Once all set points have been entered or checked, go 
to the CH 0 entrance/exit display. Press the SCROLL ENTER 
button to exit. Then press UP to go to Setup 4. 

Setup 4 GRP / DEV SETUP 

At the Setup 4, press the SCROLL ENTER button. The LCD will 
first display the Number of Channels in the first group to Average. 
All first group settings have the letter A. The second group 
settings have the letter B: 

 

If the group is set to Ø (zero), the function is disabled. The first 
group must start with Channel 1. For an example on a 16 

cylinder engine, group will be set to 16. The first 16 channels will 
be in the group. The group can be all 24 channels. If a second 
group is used, it starts with the first channel after the first group. 
Use the UP

 

and DOWN

 

buttons to select the number of channels 

that will be grouped in the averaging group. Enter Ø (zero) to 
disable this feature. Press the SCROLL

 

ENTER

 

button. Next 

enter the allowable deviation. This is a temperature adjustable 
from 0

–255 in the Units (°F or °C) chosen for the channels: 

 

If any of the grouped channels (beginning with channel 1) 
deviates from the average of the other channels by the amount 
entered, Output

 

1

 

will be energized. Use the UP

 

and DOWN

 

buttons to enter an amount. Press the SCROLL

 

ENTER

 

button. 

Next enter the enable value. This is a temperature setpoint for 
enabling the group/deviation function. If the temperature of the 
channel selected in the next setting exceeds this enabled set 
point, the group/deviation function becomes active: 

 

Use UP

 

and DOWN

 

to enter a value. Press the SCROLL

 

ENTER 

button and enter the enable channel: 

 

If this channel value is set to Ø (zero), any of the grouped 
channels will enable the group/deviation function when any of 
the channels reaches the EN

 

value. If this channel value is set 

to any other value (1-24), then ONLY that channel will enable 
the deviation function when it reaches the EN

 

value. The 

enabling channel does not have to be in the group. Use the UP

 

and DOWN

 

buttons to enter a value. 

Press SCROLL

 

ENTER,

 

and the display will go to grpb for the 

second group. If a second group is not desired, the grpb setting 
should be set to zero. With a setting of zero. pressing scroll enter 
will bring back the setup 4 display. If a second group is used, it is 
done like the first group. 

Setup 5 CH SCAN DELAY

 

At Setup 5, press SCROLL ENTER. The LCD will display the 
current display delay between channels during Normal Scan. This 
delay is also the update delay for the temperature readings. The 
channels are still scanned within 2 seconds and trip point 
comparisons are made, but the actual display of the temperature 
is delayed by this time. 

Summary of Contents for TDXM-DC Series

Page 1: ...00 02 0511 2019 07 10 Section 10 TemperatureScanner Pyrometer ModelTDXM DC Installation and Operations Manual...

Page 2: ...ly if you have any questions A visual inspection of this product for damage during shipping is recommended before installation It is your responsibility to ensure that qualified mechanical and electri...

Page 3: ...85 Serial Port 1 Modbus Integer Holding Registers 1 Replacement Parts 1 TDXM DC Mounting 2 Dimensions 2 TDXM DC Typical Wiring Diagram 3 TDXM DC I O Addresses 4 Sequence of Operations 5 Scanning 5 Set...

Page 4: ...THIS PAGE INTENTIONALLY LEFT BLANK...

Page 5: ...led Control Options Each TDXM DC model features three outputs Two outputs are the Field Effect Transistor FET and one output is the Form C Relay Each channel has three set points one for each of the o...

Page 6: ...140 mm and four mounting holes 0 25 in 6 35 mm dia see diagram 2 Align the mounting studs to the drilled holes and insert the TDXM DC from the front side of the panel 3 Place a washer and a nut on eac...

Page 7: ...input wires away from other wires 2 in 51 mm minimum For hazardous application requirements the TDXM DC complete system must be installed in accordance with the national electrical code NEC Class I D...

Page 8: ...143 Ch 18 40 018 40 048 40 072 40 096 40 120 40 144 Ch 19 40 019 40 049 40 073 40 097 40 121 40 145 Ch 20 40 020 40 050 40 074 40 098 40 122 40 146 Ch 21 40 021 40 051 40 075 40 099 40 123 40 147 Ch 2...

Page 9: ...button The display will first show The value on the right represents the channel that tripped for the SP shown From this display use the UP button to view SP2 and SP3 For each SP it is possible to vi...

Page 10: ...Y SELECTION Setup 7 RTU BAUD SETUP Setup 1 TC TYPE UNITS Each channel can be individually set to display in units C or F for Type J or K thermocouples or ignored completely When a channel is ignored i...

Page 11: ...is set to zero the function is disabled The first group must start with Channel 1 For an example on a 16 cylinder engine group will be set to 16 The first 16 channels will be in the group The group c...

Page 12: ...UP and DOWN buttons to enter the desired RTU Node Number This is adjustable from 0 255 Press the SCROLL ENTER button and enter the baud rate and number of stop bits selection 0 9600 Baud N 8 1 1 19 2...

Page 13: ...Section 10 00 02 0511 2019 07 10 9 Map of Setup...

Page 14: ...Section 10 00 02 0511 2019 07 10 10 Map of Setup continued...

Page 15: ...e following A Be sure the display is in the scan mode The unlatch operation cannot be performed while viewing the history displays B Verify that the channel in the SP3 history has cleared To do this e...

Page 16: ...w channels Error 23 Inconsistent engineering units Error 24 Channels skipped disabled Error 25 Deviation value set to 0 Error 26 Group size goes past the 24th channel CJ Error If a cold junction senso...

Page 17: ...seconds Range Type J 0 F 1538 F 0 C 837 C Type K 0 F 1999 F 0 C 1093 C Display Type Custom 7 segment 7 character type with temperature units indication and setpoint trip indication Accuracy Cold junct...

Page 18: ...Section 10 00 02 0511 2019 07 10 14 Notes...

Page 19: ...Section 10 00 02 0511 2019 07 10 15 Notes...

Page 20: ......

Reviews: