7. Error Messages and Maintenance
7.1 Error Messages
The following is a list of major error messages that can be observed. Refer
to the following table for trouble-shooting.
Error Code
Cause of Error
Action
Cleave Error
(Cleaving error are
found)
Exceeding the inspection criteria for cleave
quality
Prepare the fiber again and
retry.
Incorrect parameters setting for cleave
quality.
Check and correct the
parameters.
Over Run
(The motor detected
the overrun limit
when running
forward.)
Fiber is not loaded or not in the proper
position.
Load the fiber at the proper
position.
Inappropriate fiber program is selected.
Check and correct the
program.
Bad cleaving quality.
Prepare the fiber again and
retry.
Defects in the image processing system.
Contact service center.
Defect in the motor driving system.
Contact service center.
V-groove is dirty.
Clean the V-groove.
Focus Motor Error
Defect in the image processing system.
Contact service center.
Align Motor Error
Fiber Setting Error
Fibers are too close.
Load the fiber at the proper
position.
Low Battery
Battery has no power remaining.
See “Recharging Battery”.
SPLICE DEFECTS
See “Splicing Defects, Fusion Splicing”.
Focusing Error
(Unable to focus on
the fiber)
Fiber is dirty.
Start the splice from the
beginning. Make sure to
clean the bare portion of
the fiber.
Inappropriate fiber program is selected.
Check and correct the
program.
Incorrect parameter setting for FOCUS and
FIELD.
Check and correct the
parameters.