Maintenance
2 - 29
Checking the patient measurement current
Measurement wiring diagram
MD
Power outlet
Leakage current measuring device
Power cable
Measurement point B
Measurement
point A
RA
LF
FX-4010
LA
Conversion plug
This terminal is open.
Used for single fault condition
Patient measurement current measurement points
Measurement point A
Shorted patient cable terminal
Measurement point B
RA terminal of patient cable
Patient measurement current measurement conditions
It is necessary to measure the patient measurement current under two
conditions: normal condition and single fault condition.
Acceptance standards
Normal condition
0.01 mA or less
Single fault condition
0.05 mA or less
FX-7202
Power supply cord
Summary of Contents for CARDIMAX FX-7202
Page 10: ...viii ...
Page 12: ......
Page 18: ...Chapter 1 1 6 ...
Page 48: ......
Page 50: ......
Page 51: ......
Page 58: ......
Page 108: ......
Page 124: ......
Page 125: ...6 1 MAIN BOARD PCB 6868B component side ...
Page 126: ...6 2 MAIN BOARD PCB 6868B soldering side ...
Page 127: ...6 3 ECG AMP BOARD PCB 6867A component side ...
Page 128: ...6 4 ECG AMP BOARD PCB 6867A soldering side ...
Page 129: ...6 5 OPERATION BOARD PCB 6866C component side ...
Page 130: ...6 6 OPERATION BOARD PCB 6866C soldering side ...
Page 131: ...6 7 POWER1 BOARD PCB 6863B component side ...
Page 132: ...6 8 POWER1 BOARD PCB 6863B soldering side ...
Page 133: ...6 9 POWER2 BOARD PCB 6864B component side ...
Page 134: ...6 10 POWER2 BOARD PCB 6864B soldering side ...
Page 135: ...6 11 MARK SENSER BOARD PCB 6865 component side ...
Page 136: ...6 12 MAGAZIN SW BOARD PCB 6862 component side ...
Page 138: ......
Page 143: ...7 5 External view of recorder ...
Page 152: ......
Page 158: ...Chapter 8 8 6 ...
Page 159: ......