Maintenance
2 - 23
●
Safety inspections (6 items)
Safety-related testing methods and measurement equipment are prescribed
in safety test standards. At the actual sites where medical treatment is being
given, however, it can be extremely difficult to verify that everything meets
the standards. For this reason, when performing inspections, set up and
make acceptance determinations using the methods described below.
WARNING
If, during a safety inspection, a value that exceeds the allowable limit is
confirmed, immediately stop using the given unit; then replace the
defective part or make proper repairs. Failure to repair the unit can lead to
a serious accident.
Before checking the leakage current
There are a number of conditions for measuring leak current. For the
measured values, however, use the maximum values that are obtained.
When checking, it is also important to make comparisons with past
measured values to confirm that there are no large differences.
To measure the leakage current, use leakage current gauges (electronic
voltmeters and digital voltmeters), impedance for measurement, power
switch box, and similar equipment.
Impedance for measurement
The impedance will have a structure such as that shown below.
R1
R2
C1
V
R1 :10k
±
5% (No lead)
R2 :1k
±
1% (No lead)
C1 :0.015µF
±
5% (No lead)
V :Voltmeter
By measuring the voltage at both ends of the impedance, you can find the
leakage current.
Voltmeter
Because digital voltmeters are required to indicate true effective values for
composite waveform from DC to 1 MHz, it is recommended that you use a
voltmeter whose performance meets the prescribed standard as closely as
possible.
For use during maintenance and inspection in the field, you can use a
commercially available digital voltmeter instead. However, check the
frequency band of the voltmeter you plan to use ahead of time. To measure
a leakage current that exceeds the measurable frequency band, use it with
care paying attention to the indication on the voltmeter.
Summary of Contents for CARDIMAX FX-7202
Page 10: ...viii ...
Page 12: ......
Page 18: ...Chapter 1 1 6 ...
Page 48: ......
Page 50: ......
Page 51: ......
Page 58: ......
Page 108: ......
Page 124: ......
Page 125: ...6 1 MAIN BOARD PCB 6868B component side ...
Page 126: ...6 2 MAIN BOARD PCB 6868B soldering side ...
Page 127: ...6 3 ECG AMP BOARD PCB 6867A component side ...
Page 128: ...6 4 ECG AMP BOARD PCB 6867A soldering side ...
Page 129: ...6 5 OPERATION BOARD PCB 6866C component side ...
Page 130: ...6 6 OPERATION BOARD PCB 6866C soldering side ...
Page 131: ...6 7 POWER1 BOARD PCB 6863B component side ...
Page 132: ...6 8 POWER1 BOARD PCB 6863B soldering side ...
Page 133: ...6 9 POWER2 BOARD PCB 6864B component side ...
Page 134: ...6 10 POWER2 BOARD PCB 6864B soldering side ...
Page 135: ...6 11 MARK SENSER BOARD PCB 6865 component side ...
Page 136: ...6 12 MAGAZIN SW BOARD PCB 6862 component side ...
Page 138: ......
Page 143: ...7 5 External view of recorder ...
Page 152: ......
Page 158: ...Chapter 8 8 6 ...
Page 159: ......