41S_Rev00
71
Maintenance Menu
Diagnostic Test
The 41S has a built in diagnostic test feature that allows the operator to perform a simple one
step evaluation of splicer performance covering several different critical variables. Perform
this function in the event of splicer operation trouble.
Operation Procedure
1.
Select the Diagnostic Test icon in the [Maintenance Menu] and execute [Diagnostic
Test].The following checks will be made.
Check Item
Description
LED Check
Measures and adjusts the brightness of the illumination
LED.
Dust Check
Checks the optical path for dust or dirt and judges whether
they disturb fiber observation. If contamination exists, this
function indicates the location.
Motor Calibration Check the Motor Limit Sensor.
Arc Calibration
Automatically calibrates the arc power factor and fiber
splicing position.
I/O Port Check
Checks for normal operation of the input and output
terminals of the internal circuit.
Memory Check
Checks the memory of the internal circuit.
2.
Upon completion of all checks and adjustments, a list of results is displayed. If the
dust check result is not good, clean the objective lenses cover. In the case that
cleaning cannot eliminate contamination, there is a possibility that the contamination
may have entered the inside of the optical path. Please contact the authorized
distributor for additional instructions.
The dust check and Motor Calibration functions exist as independent instructions in
[Maintenance Menu]. It is possible to execute them independently.
Before the start of the test, remove the fibers from the splicer.
When the Motor check is completed, prepare and load the fibers into the
splicer and press SET key.