2002. Jan. Verson 3.0
131
Power Supply Facilities
(1) Voltage
100 to 115 VAC +10%
200 to 230 VAC +10%
(2) Phase
Single-phase
(3) Frequency
50 or 60 Hz
(4) Power consumption
* FLA-8000
3.0 A (100 V)
1.5 A (200 V)
* Analyzer unit Approx. 13 A (100 V)
Approx. 7 A (200 V)
* IP eraser
1.25 A (100 V)
0.7 A (200 V)
* Pictorography 12 A (100 V)
6 A (200 V)
(5) Grounding
Class D grounding. This equipment must be grounded by
connecting the power cable plug with a 2P + E AC outlet for
securing the safety, preventing external noises and providing
stable operations. The AC outlet must be grounded in Class 3
in conformity to the electric facility technical standards.
Power Supply Facilities
Summary of Contents for FLA-8000
Page 1: ...FLA 8000 OPERATION MANUAL 2002 Jan Version 3 0 Version 3...
Page 2: ......
Page 5: ...2002 Jan Verson 3 0 3 Part 1 PRECAUTIONS...
Page 9: ...2002 Jan Verson 3 0 7...
Page 18: ...2002 Jan Verson 3 0 16...
Page 19: ...2002 Jan Verson 3 0 17 Part 2 SYSTEM CONFIGURATION...
Page 23: ...2002 Jan Verson 3 0 21 Part 3 Switching On and Off...
Page 25: ...2002 Jan Verson 3 0 23 Part 4 Scanning Slide Glass Samples...
Page 50: ...2002 Jan Verson 3 0 48...
Page 51: ...2002 Jan Verson 3 0 49 Part 5 Usage of IPs...
Page 58: ...2002 Jan Verson 3 0 56...
Page 59: ...2002 Jan Verson 3 0 57 Part 6 Scanning IPs...
Page 78: ...2002 Jan Verson 3 0 76...
Page 79: ...2002 Jan Verson 3 0 77 Part 7 Scanning Gel Samples...
Page 102: ...2002 Jan Verson 3 0 100...
Page 103: ...2002 Jan Verson 3 0 101 Setting Optional Filters Part 8 Setting Optional Filters...
Page 107: ...2002 Jan Verson 3 0 105 Part 9 Daily Maintenance...
Page 109: ...2002 Jan Verson 3 0 107 Part 10 Troubleshooting...
Page 130: ...2002 Jan Verson 3 0 128...
Page 131: ...2002 Jan Verson 3 0 129 Part 11 Specifications...
Page 134: ...2002 Jan Verson 3 0 132...
Page 135: ...2002 Jan Verson 3 0 133 Part 12 Warranty...
Page 137: ...2002 Jan Verson 3 0 135...