2002. Jan. Verson 3.0
113
Code
Displayed message
Description
Countermeasures
8001H
(Warning)
FLA-8000 detected continuing
very strong signals and stopped
scanning.
Please change scanning condi-
tions and try again.
** Warning during scanning **
"PMT over-exposure detected during scanning"
All data exceed the specified over-exposure value
continuously during scanning of six lines.
Change the conditions
and retry scanning.
(Lower the sensitivity by
reducing the PMT value,
reducing the resolution
or selecting the Std.
scan mode, or increase
the area height.)
If the phenomenon does
not change, on-call, or
execute self-diagnosis.
** Door-related warning due to user's maloperation **
9201H
(Warning)
Please close the main door and
the filter door.
The sample setting door or filter exchanging door was
open. (During startup self-diagnosis or scanning)
Close the sample setting
door or filter exchanging
door.
If the phenomenon does
not change, on-call, or
execute self-diagnosis.
Countermeasures according to warning and error messages
7. When an error code is displayed, memorize it and click the OK button.
A text file having the following file name is created automatically.
xxxx Log xxxx
8. Print two text files created in steps 4 and 7. Facsimile the printout to the dealer
you purchased the FLA-8000 (dealer having engineers) or Fuji Photo Film Co., Ltd.
Serial No.
Date (Year, month, day, hour, minute, and second)
Terms
Restarting the (FLA-8000) body: Turning off the body once and then turning it on
again after several seconds.
On-call:
Calling the dealer you purchased the FLA-8000. Inform the dealer of
the following when calling it.
1. Model name of the equipment where a trouble occurs
2. Model name of the analyzer unit (DELL Optiplex GX1, for example),
as well as the name and version of the software for the FLA-8000
(FLA-8000 Image Reader V. 1.0, for example)
3. Details of the condition where the trouble occurs, displayed
message, key, and code (Example -- The trouble occurred two
minutes after switching on the FLA-8000 body. The error message
is "Self diagnosis: Connection error". Key: 4H. Code: B1H.)
Summary of Contents for FLA-8000
Page 1: ...FLA 8000 OPERATION MANUAL 2002 Jan Version 3 0 Version 3...
Page 2: ......
Page 5: ...2002 Jan Verson 3 0 3 Part 1 PRECAUTIONS...
Page 9: ...2002 Jan Verson 3 0 7...
Page 18: ...2002 Jan Verson 3 0 16...
Page 19: ...2002 Jan Verson 3 0 17 Part 2 SYSTEM CONFIGURATION...
Page 23: ...2002 Jan Verson 3 0 21 Part 3 Switching On and Off...
Page 25: ...2002 Jan Verson 3 0 23 Part 4 Scanning Slide Glass Samples...
Page 50: ...2002 Jan Verson 3 0 48...
Page 51: ...2002 Jan Verson 3 0 49 Part 5 Usage of IPs...
Page 58: ...2002 Jan Verson 3 0 56...
Page 59: ...2002 Jan Verson 3 0 57 Part 6 Scanning IPs...
Page 78: ...2002 Jan Verson 3 0 76...
Page 79: ...2002 Jan Verson 3 0 77 Part 7 Scanning Gel Samples...
Page 102: ...2002 Jan Verson 3 0 100...
Page 103: ...2002 Jan Verson 3 0 101 Setting Optional Filters Part 8 Setting Optional Filters...
Page 107: ...2002 Jan Verson 3 0 105 Part 9 Daily Maintenance...
Page 109: ...2002 Jan Verson 3 0 107 Part 10 Troubleshooting...
Page 130: ...2002 Jan Verson 3 0 128...
Page 131: ...2002 Jan Verson 3 0 129 Part 11 Specifications...
Page 134: ...2002 Jan Verson 3 0 132...
Page 135: ...2002 Jan Verson 3 0 133 Part 12 Warranty...
Page 137: ...2002 Jan Verson 3 0 135...