1-17
For Safe Operation
1
DR-ID 670 Operation Manual
897N102735
1.3.6 Precautions Related to the Load Applied to the Flat
Panel Sensor
If excessive load is applied to the fl at panel sensor, use it on a fl at fl oor or platform.
When making an exposure for the patient in a wheelchair or adjustable bed or on a stretcher, the
fl at panel sensor may be deformed (defl ection by some millimeters).
Flat panel sensor
Flat panel sensor
Avoid using the fl at panel sensor in an exposure, if the load applied to it is expected to exceed the
limits on the previous page.
In case that the fl at panel sensor is deformed, make sure that X-ray images are not adversely
affected before continuing the use of the fl at panel sensor.
The precautions below must also be observed when making an exposure.
• Do not have the patient stand on the fl at panel sensor.
• Do not place the hard devices such as spine board on the fl at panel sensor.
Excessive load is applied locally and the fl at panel sensor may be damaged.
Flat panel sensor
Spine board
Even when the fl at panel sensor is used on a fl at fl oor or platform, it may be damaged if the applied
load exceeds the limit.
Summary of Contents for DR-ID 670
Page 2: ...ii DR ID 670 Operation Manual 897N102735...
Page 24: ...1 18 For Safe Operation 1 DR ID 670 Operation Manual 897N102735...
Page 32: ...2 8 System Configuration Product Overview 2 DR ID 670 Operation Manual 897N102735...
Page 42: ...4 2 Daily Inspection and Maintenance 4 DR ID 670 Operation Manual 897N102735...
Page 44: ...A 2 Appendix A Specifications DR ID 670 Operation Manual 897N102735...
Page 54: ...Maintenance and Inspection DR ID 670 Operation Manual 897N102735...
Page 61: ...Maintenance and Inspection DR ID 670 Operation Manual 897N102735...
Page 62: ......