8.1 (7) Connection of cable to small type sensor
4.
Secure the coaxial cable with the cable clamp.
Cable clamp
G +
5.
Remove foreign matters from the terminals, and
mold the while terminal block with silicone filler.
• Cut off the tip of the silicone filler tube.
Apply silicone to the terminal block while
pressing the head of the tube against the bottom of
terminals.
At this time, care should be taken to prevent entry
of air bubbles.
Silicone filler
Outer shielding wire
Bend
Earth screw
Terminal
6.
Put the cover on the sensor.
Coaxial cable
2.
Select a mounting position on the pipe.
Note)
Mount the sensors so that the upstream and
downstream sensors can be distinguished with
each other.
Remove the cable clamp and insert the coaxial
cable through the cable lead-in port.
3.
Connect the cable to the terminal (G, +) and the
earth screw.
Note)
After connecting the outer shielding wire to
the earth screw, be sure to bend the amplifier
terminal.
Outer shielding
terminal
1.
Loosen the earth screw and the retaining knob on
the sensor using a screwdriver, then remove the
cover from the sensor.
Retaining knob
Cover
Loosen
48